| # chemlab compile layer: adding an instrument class | |
| The compile layer gives every instrument class one set of plain commands — `set-temp 80C`, `tare`, `transfer 50uL --from | |
| plate:A1 --to plate:B1` — and lowers each command to that instrument's own contract capabilities. It sits between the generic | |
| service (`chemlab/generic.py`, GenericCore) and the lab backend, plugged into the adapter hook. It has 30 classes (for example | |
| `balance`, `thermal` for temperature control and `liquid` for pipetting), a thin class and a shared layer that together cover | |
| all 203 instruments. This guide is what you need to add or extend a class: the rules, the files, the steps and the tests. | |
| ``` | |
| agent (sandbox) ── chemlab run INST VERB ... ──> GenericCore ──> adapter of INST (compile layer) ──> LabBackend ──> runtime | |
| │ generic commands unchanged: │ same agent principal, | |
| │ list/describe/read/call/wait/... │ the runtime checks again | |
| ``` | |
| An adapter only ever reaches the instrument through `core.backend` with the caller's own agent principal, exactly like | |
| `chemlab call`. Visibility rules, the output filter and the runtime's own checks stay in force. `chemlab call` stays available | |
| for everything a class does not cover. | |
| ## 1. Files | |
| | part | file | notes | | |
| |---|---|---| | |
| | framework | `chemlab/compile/{units,errors,verbs,ctx,jobs,base,fragments,shared,deck,lowering,acceptance,mapgen}.py` | shared by every class; change with care and run the full test set | | |
| | a class | `chemlab/compile/classes/<class>.py` — verb table, handlers, class checks, summary | one module per class | | |
| | its mapping | `configs/adapters/<class>.json` — one fragment per class, members and per-instrument data | one fragment per class | | |
| | host-side world | `chemlab/compile/tasks/<class>_world.py` (lab-config set-up steps), `chemlab/compile/scenes/<x>.py` (only if the runtime needs a trusted scene, like the Flex) | per class | | |
| | lowering table | `chemlab/compile/tasks/<class>_lowering.py` | per class | | |
| | acceptance tasks | `chemlab/compile/tasks/<class>_tasks.py` | per class | | |
| | tests | `tests_compile/test_class_<class>.py` (two lines, see §8) | per class | | |
| | shared hooks | `chemlab/{generic,labbackend,adapters}.py` | the generic layer; a class never needs to change them | | |
| Suites and lowering tables are found by name (`tasks/<class>_tasks.py`, `tasks/<class>_lowering.py`); fragments by directory. | |
| There is no shared registry to edit, so two classes developed at the same time never touch the same file. Do not edit `configs/all203.json`. | |
| ## 2. Steps to add a class | |
| 1. **Read** each member's agent-visible capabilities and slots (`chemlab describe INST --json`), its contract, and the | |
| manufacturer's documentation for the native vocabulary. Cite page references only; never copy manual text into code, | |
| fragments or help. | |
| 2. **Probe** each member in a small lab before writing code (what the default lab state is, what a call needs, how fast things | |
| change): `GenericLab(dir, single([...]))` and `lab.core.handle(['chemlab', 'call', ...], 'agent')`; use `class_adapters: False` | |
| in the config to see the plain generic layer. Note what needs host world (filled bath, sample, fixture) — the default lab | |
| does not provide it. | |
| 3. **Draft the mapping**: `python -m chemlab.compile.mapgen <class>` drafts a fragment from rules in `mapgen.RULES` (add a rule set | |
| for a new class there), or write the fragment by hand. `python -m chemlab.compile.mapgen <class> --compare` shows how much the | |
| rules find for an existing fragment. | |
| 4. **Write the class module** (§3) and the **fragment** (§4). Keep instrument facts in the fragment, class logic in the module. | |
| 5. **Lowering table** (§7): per member, verbs in order with the native calls each must make, written from the contracts. | |
| `python -m chemlab.compile.lowering --class <class>` must PASS (calls equal, twin state equal after every step, one path). | |
| 6. **Acceptance tasks** (§7): 8–10 for classes with 5 or more members (4–6 for small classes), at least two recovery tasks, | |
| gold with `chemlab run` passes in the sandbox, NoAct fails: `python -m chemlab.compile.acceptance --class <class>`. | |
| 7. **Tests**: `tests_compile/test_class_<class>.py`; then the full set (§8), including the generic parity for your members. | |
| 8. **Document** in the fragment: membership decisions with reasons (who is in, who is out and why, primary / secondary) and | |
| verbs not supported per member with reasons. Fix runtime or contract problems in the runtime or contract, not in the class. | |
| ## 3. The class module | |
| ```python | |
| from ..base import ClassAdapter, finite | |
| from ..errors import invalid, refused | |
| from ..verbs import Arg, Opt, Verb, save_to_opt | |
| class CentrifugeAdapter(ClassAdapter): | |
| CLASS = 'agitation' # = the fragment's "class" | |
| TITLE = 'spin, shake, mix' | |
| ABOUT = 'one paragraph for describe' | |
| STANDARDS = 'SiLA 2 ..., PyLabRobot ..., ASTM / pharmacopoeia terms' | |
| ALWAYS = ('status',) # verbs every member has without a mapping entry | |
| UNIT_KINDS = ('rotation', 'duration') | |
| PATH_GROUPS = (('start', 'stop', 'result'),) # one native path per group (§5.2) | |
| CODES = {'LID_OPEN': ('LID_OPEN', 'the lid of {inst} is open.', 'close it: `chemlab run {inst} close lid`.')} | |
| def build_verbs(self): | |
| return {'spin': Verb('spin', 'spin at a speed for a time', self.h_spin, | |
| positionals=[Arg('SPEED', 'e.g. 4000rpm', kind='rotation')], | |
| options={'time': Opt('run time', kind='duration', metavar='TIME', default='5min')}, | |
| preconditions=['lid closed', 'speed within the rotor range'], | |
| errors=['OUT_OF_RANGE (usage)', 'LID_OPEN (refused)'], | |
| reference='manual pages ...', example='chemlab run INST spin 4000rpm --time 5min')} | |
| def summary(self, values): # values: {slot: value} or a Ctx (use self.slot_si / self.val helpers) | |
| return ['...one line...'], {'json': 'view'} | |
| def h_spin(self, ctx, p): | |
| m = self.vmap('spin') # this member's mapping for the verb | |
| w = p.pos[0] # already SI (rad/s): verbs parse units | |
| lo, hi = self.arg_range_si(m['cap'], m['arg']) # range from the contract, in SI | |
| if (lo is not None and w < lo) or (hi is not None and w > hi): | |
| ctx.fail(invalid('OUT_OF_RANGE', '...', 'give a speed from ... to ...')) | |
| self.requirements(ctx, 'spin') # the member's "requires" entries for this verb | |
| ctx.check(not ctx.v(self.slot('lid_open')), 'LID_OPEN', 'the lid is open.', 'close it first.') | |
| ctx.call(m['cap'], {m['arg']: self.arg_value(m['cap'], m['arg'], w)}, 'set the speed to %.0f rpm' % (w * 9.549)) | |
| return {'text': 'ok: ...\n', 'json': {...}} | |
| ``` | |
| What the base gives every class (do not re-implement): argument and unit parsing, `--dry-run` (checks run, every native call is | |
| listed, none is sent), destructive verbs (`destructive=True`) only preview without `--yes`, `--save-to`, unknown verbs with close | |
| matches, unsupported verbs with reasons (exit 6), `describe` / `help INST VERB` / `run INST VERB --help` / `run INST`, `read` in | |
| friendly units with the class view on top, error format, partial-progress reports, refusal explanation, the mapping check | |
| against the contract, `status`. | |
| The request context `ctx` (`ctx.py`): `ctx.v(slot)` (current value; `slot#key` for a key of an object slot via `self.slot_si`), | |
| `ctx.call(cap, args, step, refusal=, readback=, query=)` (one native call, validated against the contract schema first, recorded, | |
| read back), `ctx.check(ok, CODE, message, fix)` (a class precondition), `ctx.fail(err)` (raise with the steps done), | |
| `ctx.advance(seconds, until=fn, poll_s=)` (let lab time pass; the only way a class verb may move time — only in verbs that say | |
| they wait; the generic `chemlab wait` takes 60, 30s, 5min or 1.5h), `ctx.plan` / `ctx.notes` / `ctx.done`. | |
| ## 4. The fragment (`configs/adapters/<class>.json`) | |
| ```json | |
| {"class": "thermal", "adapter": "chemlab.compile.classes.thermal:ThermalAdapter", "about": "...", | |
| "membership": {"primary": [...], "secondary": {...}, "excluded": {"slug": "reason"}}, | |
| "members": { | |
| "julabo_corio_cd200f": { | |
| "role": "primary", | |
| "verbs": {"set-temp": {"cap": "thermal.set_temperature", "arg": "temperature_K"}, | |
| "heat": {"cap": "circulation.set_running", "arg": "running"}, | |
| "stir": {"speed_cap": "stirring.set_speed", "speed_arg": "speed_rad_s", "on_cap": "stirring.set", "on_arg": "on"}, | |
| "safety-limit": {"cap": "safety.configure", "arg": "limit_K", "args": {"mode": "limit"}}}, | |
| "slots": {"temp": "thermal.indicated_temperature_k", "set": "setpoint.fixed#temperature_k", | |
| "display": {"slot": "display.temperature", "unit_slot": "unit"}}, | |
| "requires": {"set-temp": [{"slot": "interface.remote", "value": true, "code": "REMOTE_OFF", | |
| "message": "{inst} takes set points only in remote mode.", "fix": "`chemlab run {inst} remote on`."}]}, | |
| "codes": {"circulation.set_running:PRECONDITION_FAILED": {"code": "NOT_READY", "message": "...", "fix": "..."}, "UNSTABLE": {...}}, | |
| "unsupported": {"stir": "no stirrer: the bath is mixed by the circulation pump"}, | |
| "alternatives": {"stir": "what to do instead (shown as the fix)"}, | |
| "targets": {"zone": {"arg": "zone_index", "values": {"1": 0, "2": 1}}}, | |
| "locations": {"tray_01": {"position": 1}}, | |
| "path": "acquisition", "path_exceptions": ["sieving.measure"], | |
| "params": {"tolerance_k": 0.5}, | |
| "notes": {"heat": ["shown in help INST heat"]}}}} | |
| ``` | |
| Keys a member may use: `verbs.<verb>` with `cap`, `arg`, `args` (fixed arguments), `<name>_cap` / `<name>_arg` for multi-call verbs, | |
| `reads` (state slots, for read-only verbs); `slots.<role>` (slot id, `slot#key`, or `{"slot", "unit", "unit_slot"}`); | |
| `requires`, `codes` (runtime code or `cap:CODE` -> class code, message, fix), `unsupported` (verb -> reason; every class verb a | |
| member lacks needs one), `alternatives`, `hides` (shared verb -> reason, §5.3), `shared_fallthrough`, `targets`, `locations`, | |
| `path` / `path_exceptions`, `params`, `notes` (INST is replaced by the instrument id), `role`, and `adapter` (a member-specific | |
| subclass, rarely needed). | |
| `requires` tests (base `requirements`; several tests in one entry must all hold; `"slot"` may be `slot#key`, or give `"key"`): | |
| `value`, `not` (list), `in` (list), `prefix` (text or list of texts), `contains_not` (an item a list / text slot must not | |
| contain), `min` / `max` (inclusive) / `gt` / `lt` (numbers; a missing value fails), `truthy`, `falsy`, `null`, `set` (not null); an | |
| entry without a test asks for a truthy value. Message and fix may use `{inst}` and `{value}`. | |
| Slot references in a mapping: `check_mapping` checks the values of keys named `slot`, `reads`, `slots_read` or ending in `_slot` | |
| when they look like slot ids (`component.field`, optionally `#key`, or a visible slot id). Other values under such keys (a rack | |
| slot `"A1"`, a number) are class data and are not checked. Never copy ranges, enums or units into a fragment: the adapter reads them | |
| from the contract (`arg_range_si`, `arg_enum`, `arg_unit`). `check_mapping` verifies every capability (agent-visible and | |
| implemented), argument name and slot against the contract at first use and in the tests. | |
| Members are matched by asset type (`asset_type_id` without `instrument.`), so a lab may name its instruments freely ("flex", | |
| "bal1", "bal2"). An entry that names its own `"adapter"` keeps it. The lab config switch `"class_adapters"`: `false` (none), a | |
| directory path (other fragments), or a list of class names. | |
| ## 5. Rules | |
| ### 5.1 Units: convert to the unit each argument declares | |
| The 203 contracts declare 145 unit strings; many arguments are not SI (°C / degC on 78 arguments, g, min, rpm, mL, mbar, nm, h, | |
| days, %), and one unit is spelled several ways (m^3 / m3, °C / degC, 1/s / s^-1, ohm / Ω). `units.py` has the alias table | |
| (`ALIASES`, `DECLARED` / `_DECL`: unit -> kind, factor, offset). Rules: | |
| - Verbs parse what the agent types into SI (`Arg(kind='temperature')`, `Opt(kind='duration')`; bare numbers take the kind's default | |
| unit: µL, °C, s, rpm, mbar, g). Kinds: volume, mass, length, wavelength, wavenumber (cm-1), duration (fs ... days), temperature, | |
| temperature_delta, heating_rate, rotation, angle, pressure, flow, mass_flow, fraction, frequency (Hz ... GHz), frequency_offset | |
| (signed Hz, e.g. NMR), count_rate (/s, kcps), magnetic_field (T, mT, G), power, voltage, current, force, energy, speed, area, | |
| concentration (mol/L, mM, µM, nM, pM), density (also mass concentration: g/L, mg/L, µg/L, mg/mL, µg/m3), amount (mol, mmol, µmol, | |
| nmol, pmol), conductivity. Flow also takes m3/h, L/h and standard flows (sccm, slpm / slm, sccs), fraction | |
| ppm / ppb, force pN / nN / µN / mN / kN, energy eV / keV, power µW / mW / µcal/s / cal/s, duration us / µs / μs; the declared | |
| unit `1/min` is rpm (Linari). Kinds field_rate (mT/s), force_rate (N/s), resistance (Ω, kΩ, MΩ), sheet_resistance (Ω/sq), | |
| moment (emu, A·m²), heat_capacity (J/K) and number (plain) are framework kinds too. Add a missing unit to units.py (the shared | |
| table), not as a local copy in a class. | |
| - `units.fmt` never shows a non-zero value as 0: it moves to a smaller unit (1 Pa -> `1 Pa`, 5e-12 N -> `5 pN`, 2e-6 -> `2 ppm`; | |
| durations under 0.05 s -> ms / µs; pressures below 1 mbar keep mbar with 3 significant digits, 5 Pa -> `0.05 mbar`; | |
| kg/m³ below 100 shows as g/L, mg/L or µg/L, above as g/cm³; X-ray energies in keV / eV). `read` never shows a K slot that | |
| holds a temperature difference (id or description says diff / delta / tdiff / 温差) in °C. Keep 15 significant digits when you round a native argument (`units.sig(x)`), never | |
| `round(x, n)` (0.02° is 3.49e-4 rad; `round(.., 8)` breaks the runtime's grid check). | |
| - Handlers convert SI into the argument's declared unit with `self.arg_value(cap, arg, si)` (or `units.parse_for(text, schema)` | |
| in one step), and check ranges in SI with `self.arg_range_si(cap, arg)`. Never assume an argument is SI. | |
| - Slots: `self.slot_si(values, spec)` gives SI for a slot in any declared unit, or a display slot whose unit is in another slot | |
| (`{"slot": "display.temperature", "unit_slot": "unit"}` for BINDER C/F). | |
| - A temperature difference is `temperature_delta` (0.5K == 0.5C), not a temperature. Gauge and absolute pressure are never | |
| converted into each other: the argument name says which one it is. | |
| - Tests on real arguments of several units are in `tests_compile/test_units.py` (`RealArguments`); add your class's odd units there | |
| or in your class's own test file. | |
| ### 5.2 One native path per verb | |
| Some instruments have two native paths for the same action (DMA 5002 / Abbemat `measurement.start` vs `acquisition.start`, HX204, | |
| MP70, Selekt, Extrahera). A member's mapping picks one path, and the verbs of one group (`PATH_GROUPS`, e.g. start / stop / | |
| result) use that path: their capabilities share one namespace (`"path"`, or the namespace of the first capability), and a result | |
| read from slots reads slots that the start capability changes (its contract `state_effects`). Exceptions are explicit | |
| (`"path_exceptions"`, which may also list state slots a result reads although the start capability's `state_effects` leave them | |
| out, e.g. the 8697 `sequence.start`). Enforced statically by `check_paths` (in `check_mapping`) and dynamically by the lowering check | |
| (`path_problems`). Prefer the acquisition core where both exist unless the manual path is the one the class | |
| verbs mean. | |
| ### 5.3 Shared vocabulary, the shared layer and the override rule | |
| `compile/shared.py` defines the shared verbs as factories and `SharedAdapter` (class `shared`): | |
| | verb | lowers to (mapping key) | | |
| |---|---| | |
| | `power on\|off` | `power`: `{"cap": "power.set", "arg": "enabled"}` or a toggle `{"cap": "power.toggle", "toggle": true}` + `slots.power` | | |
| | `status` | the class view (every adapter) | | |
| | `open PART` / `close PART` | `open`: `{"parts": {"door": {"cap": "door.set", "arg": "angle_rad", "open": 1.57, "close": 0.0}, "lid": {"open_cap": "lid.open", "close_cap": "lid.close"}}}` | | |
| | `ack` / `estop` | `{"cap": "alarm.acknowledge"}` / `{"cap": "emergency.stop"}` | | |
| | `method KEY=VALUE ...` | `{"cap": "method.configure", "keys": {...class dictionary...}, "passthrough": true}` (§5.5) | | |
| | `start` | `{"cap": "acquisition.start", "job": "operation" \| "state" \| "none", "state": {...}}` (§5.4) | | |
| | `stop` / `result [--save-to F]` | `{"cap": ...}` / `{"cap": ...}` or `{"reads": [...]}` | | |
| | `wait [--max T]` | the caller's latest job on the instrument | | |
| `compile/shared_gen.py` generates `configs/adapters/shared.json` with `"layer": "shared"` for 198 instruments (`python -m chemlab.compile.shared_gen | |
| --write`; never hand-edit it — `test_shared_all203` checks that the generator reproduces it): power 194, open/close 101 (138 named | |
| parts), method 127, start 126, stop 118, result 108, ack 39, estop 21. The other five are the thin class (`configs/adapters/ | |
| thin.json`, `classes/thin.py`: oxitopi, puresolvmd, dmt152, Liberty Blue, oligosynt), reviewed by hand. If your class needs a | |
| verb name the shared layer also has (power, open, close, ack, estop, method, start, stop, result, wait), your class verb wins on | |
| your members; give it the same meaning. In `method`, a native argument name with a bare number takes the unit the contract | |
| declares for it (`duration_s=60`); with a unit it is converted (`duration_s=1min`); class keys always take units. **Override rule:** an | |
| instrument in a class fragment and in the shared fragment gets a CompositeAdapter: class fragments first (primary role before | |
| secondary), the shared layer last; a verb belongs to the first part that supports it. So a class verb overrides the shared one, | |
| and the shared layer still answers the verbs the class does not have (`tests_compile/test_shared.py`). **A class's "unsupported" | |
| also wins over the shared layer:** a verb in your class table that a member marks unsupported (with its reason and | |
| alternative) is answered by your class with exit 6 on that member, never by the shared layer — unless the member lists it in | |
| `"shared_fallthrough": ["open"]`, which hands that verb back to the shared layer (e.g. a class that has no part table of its own | |
| for `open`). **Hiding a shared verb:** a class that replaces a shared verb by its own (GPC `run` instead of `start`, TLC | |
| `chromatogram` instead of `result`) declares it: class attribute `HIDES = {"start": "use `chemlab run INST run`"}` (all members) or | |
| member `"hides": {"result": "..."}`; or name the replacing verb: class `COVERS = {"start": "measure"}` / member `"covers": | |
| {"start": "measure"}` (reason: use `chemlab run INST measure`; it applies only on members where the covering verb is supported); the hidden verb answers exit 6 with that reason on those members, `run INST` lists it under | |
| "not on this instrument", `help INST start` says what replaces it; `shared_fallthrough` undoes it per member. | |
| **Ownership order on a composite instrument** (fragments.py `CompositeAdapter.owner`): (1) the first class part that supports | |
| the verb (primary before secondary); (2) else a class part that marks it unsupported or hides it (its reason); (3) else the | |
| shared layer. A class's "unsupported" therefore never blocks another class: on KS 4000 / Elma (agitation primary + thermal | |
| secondary) agitation's `"unsupported": {"set-temp": ...}` leaves `set-temp` to thermal. Class modules may reuse the factories | |
| (`self.table.update(shared.base_verbs(self))`). | |
| **Shared `method` sends the full argument set** (most contracts require every argument of `method.configure`): the keys the | |
| agent gives, the mapping's fixed `args`, and for every other required argument (and every argument chemlab sent last time) in | |
| this order: what the instrument reports (a visible slot `<namespace>.<arg>` such as `method.mode`, or a key of a visible object | |
| slot holding several arguments such as `method.current`), the last method chemlab sent on that capability (request log, any verb | |
| or `chemlab call`), the contract's `example_arguments`. The reply names the source of every value it did not get from the agent | |
| (`--json` `sources`); a required argument with no source is a usage error naming it; values outside the contract ranges are a | |
| usage error (BAD_SETTING) before anything is sent. Helpers for classes: `shared.full_method_args(ad, ctx, cap, given, fixed)`, | |
| `shared.reported_method`, `shared.last_sent`, `shared.method_sources_text`. The other generated shared verbs never need | |
| arguments they do not supply (checked over all 198 members). | |
| ### 5.4 Jobs: three kinds of long action | |
| | kind | when | class | status / wait / cancel | | |
| |---|---|---|---| | |
| | operation | an async capability returns an operation_id (most `*.start`) | `jobs.OperationJob` | poll / cancel of the session | | |
| | run-type / state | progress only in slots (Flex protocol runs, furnace programmes, GC `run.start`, `cycle.start`) | `jobs.StateJob` (or a subclass, e.g. the liquid class's `RunJob`) | slot values; cancel = the class's stop call | | |
| | timed state change | no run object (reach a set point, a stable reading) | `jobs.ConditionJob` | a condition on the state; cancel stops waiting (or `on_cancel`) | | |
| A verb that starts a long action returns `job-N` (text and `--json` `job`). The agent follows it with the generic `chemlab status | |
| job-N`, `chemlab wait job-N [--max T]`, `chemlab cancel job-N`; `chemlab status` lists the caller's jobs. Jobs belong to their | |
| caller. In `--json` the job's `status` is the framework's (running / succeeded / failed / cancelled); the class's own word is in | |
| `<kind>_status` (e.g. `run_status: completed`). Job states: `jobs.RUNNING` lists the known running words (incl. `injecting`, | |
| `equilibrating`, `run`, `post`, ...), and any other text state that is not terminal (`jobs.TERMINAL`: completed, succeeded, | |
| failed, cancelled, aborted, stopped, idle, ready, ...) also counts as running, so an unknown run phase keeps the job alive. An | |
| operation whose poll says `succeeded` while its result names a known running phase (`phase: injecting`) is still running. | |
| `OperationJob(..., poll_s=N)` (or the member's `"params": {"job_poll_s": N}`) sets the lab seconds between polls while waiting: | |
| use 10-60 s where the runtime answers a poll with a whole curve. `self.wait_until(ctx, test, what, max_s, poll_s)` (with | |
| `self.moving_done([slots])`) waits for a lid, sash, drawer or stage to stop moving; refused NOT_SETTLED when it does not. Blocking variants (`wait-temp`, `weigh --stable`, `set-temp --wait`) move lab | |
| time with `ctx.advance` and say how long they waited. | |
| ### 5.5 Method keys per class | |
| `method` is the hardest verb (chromatography alone has 9 method shapes). Each class defines a small dictionary of canonical keys | |
| with units — e.g. chromatography `flow=1mL/min oven=40C..250C@10K/min inj=1uL wavelength=254nm`, spectroscopy `from=200nm to=800nm | |
| step=1nm` — and each member maps them: `"method": {"cap": ..., "keys": {"flow": {"arg": "flow_ml_min", "kind": "flow"}}}`. Keys the | |
| dictionary does not have pass through by native argument name (`"passthrough": true`), still parsed into the declared unit. | |
| Document the dictionary in the class module docstring and in `help INST method`. | |
| ### 5.6 Named locations and `--zone / --channel / --cell / --head` | |
| - A deck of labware (Flex): `deck.py` (`LABWARE:WELL`, lists, ranges in Opentrons order, trash, home), built from the lab-config | |
| `"deck"` plus the contract's slot geometry; the trusted host's scene registers the same deck with the runtime. | |
| - Other named places (autosampler trays, rack positions, robot ports): `self.location(ctx, name)` looks the name up in the lab | |
| config entry's `"locations"` (task / lab specific) and then the mapping's `"locations"` (fixed by the instrument), with close | |
| matches on a miss. | |
| - Several identical units on one instrument: one option each, `--zone` (heating blocks, reactor pads), `--channel` (pumps, gas | |
| lines), `--cell` (measuring cells), `--head` (measuring / dosing heads): `verbs.target_opt('zone')` in the verb and the member's | |
| `"targets": {"zone": {"arg": NATIVE, "values": {"1": native, ...}, "required": false}}`; `self.target_args(ctx, parsed)` lowers | |
| them. Do not invent per-class spellings (`--pad`, `--line`, ...). | |
| ### 5.7 Records: `--save-to FILE` | |
| Verbs that return a record (results, spectra, chromatograms, images, a temperature log) take `--save-to FILE` (`save_to_opt()`): | |
| the full reply is written as JSON into the agent's work dir (paths outside it are a usage error, nothing sent). Files the runtime | |
| writes are already copied into the work dir by the generic layer. | |
| ### 5.8 Errors and refusals | |
| Every error says what is wrong, how to fix it, whether the instrument changed and the steps done before it (composite verbs): | |
| ``` | |
| error[refused/UNSTABLE]: the reading of ohaus_px224e is not stable (a draft-shield door is open), so it cannot tare. | |
| fix: close the doors (`chemlab run ohaus_px224e door close`), then retry with --wait 30s. | |
| instrument changed: no | |
| request: r-0001 | |
| ``` | |
| Exit codes are the generic ones: 2 usage (nothing sent), 3 refused, 4 unexpected, 6 not supported on this instrument. Check what | |
| you can before sending (class checks with `ctx.check`, member requirements in `requires`); when the runtime still refuses with a | |
| generic code, `explain_refusal` names the reason from the member's `codes`, the class `CODES`, or the contract preconditions with | |
| the state already read (`It checks ...; now ...`). Override `explain_refusal` only to infer a reason from the state (the balance | |
| class turns the XPR's bare PRECONDITION_FAILED into UNSTABLE / POWER_OFF / ZERO_RANGE). | |
| ### 5.9 World set-up belongs to the host | |
| What the lab operator does before the agent starts (fill a bath, register a sample, install a fixture, place a load) is | |
| lab-config set-up (`"setup"` steps: capability calls as operator or `"evaluator": true`, `{"host": method, "kwargs": ...}`) or task | |
| `world(lab)` code — never class verbs and never adapter code. Examples: `tasks/thermal_world.py` (JULABO bath model before power-on, | |
| Elma tank fill), `tasks/balance_tasks.py` (`place`: pan.set_load as the evaluator, or the XPR cell's contact force). A runtime that | |
| needs a trusted scene during the session (the Flex's tip confirmations) gets a scene class in the entry (`"scene": | |
| "module:Class"`, `scenes/flex.py`). | |
| **The default lab.** `configs/all203.json` is a ready lab: `chemlab/default_world.py` (generated by | |
| `python -m chemlab.default_world_gen` from the class worlds: lowering `config()`s and, recorded as steps, the function worlds of | |
| particle_surface and quant_analyzer) + `chemlab/default_world_overrides.py` (hand, with sources), merged by make_config. Its steps | |
| carry `"default": GROUP` (utility, accessory, consumable, standard, state). A class world that needs a different start sets | |
| `"default_world": false` on the entry (all default steps off, e.g. "_empty" variants, `W.bare`) or adds a step | |
| `{"replace_default": [GROUP, ...]}`; a world step identical to a default step runs once. The lab's containers (bottles, standards, | |
| consumables, cylinders) are in `configs/inventory_default.json` (`chemlab/inventory.py`; `chemlab list --inventory [ID]`). After a | |
| class world changes: re-run the generator, then make_config, then `tests_generic/test_default_world.py`. | |
| ## 6. Pitfalls | |
| - The default lab is a ready world (see 5.9): a task that tests a missing fixture or an empty cell must switch the default off | |
| for that entry. Probe first; put the world in | |
| tasks, the reason in `codes`. | |
| - Lab time moves only with `chemlab wait` (or a verb that says it waits). Clock steps are applied lazily per instrument; a read | |
| flushes them. A changed load or door shows up only after a clock step. | |
| - The twin's own "stable" / "reached" flags can rise early (an older XPR twin said stable 1.5 s after a load change while 0.4 % low). | |
| Waiting verbs add a settle check (two readings 1 s apart within 2 d) instead of trusting one flag. | |
| - Member-specific rules hide in the contract preconditions and the runtime (JULABO only takes set points in remote mode; KS 4000 | |
| takes safety limits only with heating off; CellXpert set points in 0.1 °C steps; the SPX zero range counts from the calibration | |
| zero). Turn each into a `requires` entry or a class check with a fix. | |
| - Composite verbs report partial progress: do the checks that can fail before the first native call (e.g. stir checks the | |
| fixture before setting the speed). | |
| - Never call host methods from an adapter, never use the evaluator principal in a class, never put manual text into code or docs. | |
| - A class module that fails to build its adapter (an exception in `__init__` / `build_verbs`) is left out of the lab and recorded | |
| in `fragments.LOAD_ERRORS`, so other classes keep working; the class tests build adapters with `strict=True` and fail on it. | |
| - A handler may raise the generic `usage(...)` (e.g. from `coerce`): it stays a usage error (exit 2), not INTERNAL. | |
| - Write fragment files atomically (`fragments.write_json_atomic(path, obj)`: temp file + rename) from your generator: a lab | |
| loading fragments while you rewrite one otherwise reads half a file (the loader retries once). | |
| - Runtimes with no advance mode whose timed events run only in `host_update()` (Agilent 8697; DR6000 also syncs on access) get it | |
| called after every clock step by the lab backend (`inst.clock_hook`); no scene is needed for that any more. | |
| ## 7. Lowering table and acceptance tasks | |
| `tasks/<class>_lowering.py`: | |
| ```python | |
| from .thermal_world import config # optional: the lab config with the class's world (else all203) | |
| CASES = {'julabo_corio_cd200f': {'prep': None, 'steps': [ | |
| ('set-temp 40C', []), # refused by a class check: no native call | |
| ('remote on', [('interface.set_remote', {'enabled': True})]), | |
| ('set-temp 40C', [('thermal.set_temperature', {'temperature_K': 313.15})]), | |
| ('#wait 2', None), # lab time passes on both labs | |
| ('wait-temp --max 30min', []), | |
| ]}} | |
| ``` | |
| Expected calls are written from the contracts (capability, arguments in the declared unit), not derived from the mapping. For | |
| analysis steps whose arguments are measured arrays, an expected item may be `(capability, callable(arguments) -> bool)`, or the | |
| whole expectation a `callable(calls) -> bool`; inside expected arguments the matchers `lowering.Measured()` (a finite number or | |
| a non-empty list / object) and `lowering.Obs(n, fixed, measured)` (n observation objects: fixed fields equal, measured fields | |
| present) stand for measured values; `None` means not checked (use it only for state-dependent calls). `('#host what', fn)` runs | |
| `fn(backend)` on the lab and on the twin (the lab operator swaps a sample in the middle of a case) and compares their states. The | |
| check runs each verb on a lab with adapters, compares the native calls, replays the same calls (same request ids) and clock | |
| advances on a twin `LabBackend` without chemlab, and compares the instrument's full evaluator view after every step; it also | |
| checks the one-path rule (on a composite instrument, each part only on the verbs it owns). Cover every verb of every member, refusals included (`[]`), a dry run, and an unsupported verb. | |
| `tasks/<class>_tasks.py`: `TASKS = [{'id', 'kind', 'instruments', 'config' (fn(cfg) -> cfg, optional) or 'lab_config' (fn() -> | |
| whole config), 'world' (fn(lab), optional), 'instruction', 'gold' (bash with `chemlab run`), 'check' (fn(c))}]` and | |
| `main(only, noact)` returning `run_suite(...)`. Checkers read `c.ev(iid)` (evaluator view, truth slots included), `c.work_file()`, | |
| `c.number_file()`; `c.near()`, `c.equal()`, `c.add()`. Gold must pass in the sandbox, NoAct must fail; include recovery tasks (the | |
| gold triggers the error, reads `--json` `.error.code`, and recovers as the fix says). Results go to `compile_results/<class>/`. | |
| ## 8. Tests (all must pass before reporting) | |
| ```bash | |
| cd Chemistry-Sim/chemlab/tests_compile | |
| PY=python3 # Python 3.10+ with the runtime requirements (numpy < 2.4, scipy, pillow, ...) | |
| $PY -m unittest test_units test_framework test_shared test_visibility # framework + visibility of every adapter output | |
| $PY -m unittest test_class_<class> # mapping fits every contract, lowering, acceptance | |
| CHEMLAB_PARITY_CLASSES=<class> $PY -m unittest test_generic_parity test_visibility # generic sweeps + visibility on your members | |
| cd ../tests_generic && $PY -m unittest test_generic_core test_generic_sandbox test_mcp_shell # generic layer unchanged | |
| cd .. && $PY -m unittest discover -s tests # the Flex pilot's gate and sandbox | |
| ``` | |
| `test_class_<class>.py` is two lines: | |
| ```python | |
| from helpers import class_suite | |
| ThermalClass = class_suite('thermal') | |
| ``` | |
| `test_generic_parity` runs `sweep.equivalence` and `sweep.ready` (with the visibility scan of describe / read output, now with | |
| class sections) on the members of the class fragments — `CHEMLAB_PARITY_CLASSES=a,b` limits it to your classes — each run in its | |
| own directory `build/compile_parity/<pid>-<time>/` (generic_results/ untouched), so runs at the same time never collide. | |
| `test_visibility` takes the same filter. Lowering runs go to `build/lowering_runs/<class>/<instrument>/`. A mapping that does not | |
| fit its contract never breaks the generic commands of that instrument (they fall back to the generic path; `chemlab run` answers | |
| MAPPING_ERROR), but `test_class_<class>` fails on it. | |
| ## 9. Coverage and examples | |
| All 203 instruments have `chemlab run` verbs: 198 have a primary class (3 of them also a secondary class), 5 are in the thin | |
| class (shared verbs only, reviewed by hand) and the shared layer adds the common verbs wherever a contract has them. The 30 | |
| classes are agitation, balance, bioassay, chromatography, coating, digestion, dosing, electrical, electrochem, enclosure, | |
| evaporation, extraction, formulation, gas_fluid, imaging, labware_automation, liquid, mechanical, particle_surface, | |
| property_meter, quant_analyzer, reactor, sample_intro, solid_prep, spectroscopy, structure, thermal, thermal_analysis, | |
| titration and wash_sterilize (one fragment each in `configs/adapters/`). | |
| Small classes to read first: `balance` (3 members: power, zero, tare, weigh [--stable], unit, door, status) and `thermal` | |
| (set-temp [--wait], heat, temp, wait-temp, stir, safety-limit, remote, power, status). A member may bring its own adapter class | |
| (`"adapter"` in its fragment entry), as the NE-1000 syringe pump does inside `liquid` (`classes/syringe_pump.py`). | |
| Twins refuse to stop or change another caller's run (ACCESS_DENIED): the classes say NOT_YOURS, and their BUSY fix says to | |
| wait instead of stop when the run is not the caller's job. | |